Skip to main content

+1 437-505-5995 sales@chromkit.ca

Buying guide

Choosing a flash cartridge, and why many bad runs are decided before you start

By Dr. Nazmul Alam, analytical chemist · August 2, 2026 · Updated October 4, 2026

Short answer

Choose the flash cartridge from your TLC plate. Biotage suggests a target Rf between 0.15 and 0.4. Turn Rf into column volumes (CV = 1/Rf), and use the gap in CV to the nearest impurity to decide the load: SiliCycle's chart goes from 1% of the silica mass for a close pair to 10% for a wide one. Dry load if the sample only dissolves in a strong solvent.

A lot of flash runs that fail were already lost before the pump started. The cartridge was the wrong size, or the sample went on in the wrong solvent. A gradient can’t fix either one afterwards.

Here’s how I pick, and what I check first when a separation won’t resolve.

Start at the TLC plate

Run the TLC in the solvent system you plan to use on the column. Biotage suggests adjusting the solvent so the target compound has an Rf between 0.15 and 0.4.

Rf turns into column volumes very simply. The number of column volumes (CV) a compound needs to elute is about 1/Rf:

RfColumn volumes to elute
0.52.0
0.42.5
0.33.3
0.254.0
0.25.0
0.156.7
0.110

Too high an Rf and everything comes out close together near the start. Too low and the compound takes many column volumes, the band spreads, and you use a lot of solvent.

The gap to the nearest impurity decides the rest

Biotage measures selectivity as ΔCV: the difference in column volumes between your compound and the nearest impurity. The bigger it is, the easier the separation.

Example: your compound is at Rf 0.3 (3.3 CV) and the nearest impurity is at Rf 0.4 (2.5 CV). ΔCV is 0.8.

If ΔCV is close to zero, a bigger or more expensive cartridge won’t help. Biotage’s advice is to try solvents with different selectivity first, then adjust the strength. Fix the solvent system before you spend money on the column.

How much to load

SiliCycle publishes a loading chart for its cartridges based on ΔCV. It works out to about 1% of the silica mass for a close pair, up to 10% for a wide one:

CartridgeΔCV 0.1 to 0.6ΔCV 0.7 to 1.2ΔCV 2.5 to 3.1ΔCV over 6
4 g40 mg80 mg200 mg400 mg
12 g120 mg240 mg600 mg1.2 g
25 g250 mg500 mg1.25 g2.5 g
40 g400 mg800 mg2.0 g4.0 g
80 g800 mg1.6 g4.0 g8.0 g
120 g1.2 g2.4 g6.0 g12 g

So in the example above (ΔCV 0.8), a 12 g cartridge takes about 240 mg. The same cartridge takes 5 times more of a mixture with ΔCV around 3.

The mistake I see is sizing the cartridge on sample mass alone and overloading a hard separation. The band broadens, it tails, and the fractions need running again. The cartridge did nothing wrong.

These are a supplier’s numbers, so treat them as a starting point and adjust from your own runs.

Irregular or spherical silica

SiliCycle sells its flash cartridges in two silica grades: irregular 40 to 63 µm, and spherical 25 µm in the Premium line.

Irregular 40 to 63 µm is the usual choice, and I’d use it for most runs. It’s also the same silica as their TLC plates, so the plate and the column match.

Spherical 25 µm costs more. In general, smaller and more uniform particles give sharper bands, at higher back pressure. I’d consider it when ΔCV is small, when you have to load heavily, or when the compound is valuable enough that a second run costs more than the cartridge.

Normal phase or reversed phase

Biotage’s rule of thumb is about solubility. If the crude mixture dissolves best in a non-polar solvent, use normal phase on silica. If it dissolves best in a polar solvent, reversed phase (C18) is a better fit. Peptides are usually purified by reversed phase.

One practical difference: Biotage says reversed-phase flash methods are usually developed on HPLC, not TLC. So the TLC-based sizing above is for normal phase.

The loading mistake, and it’s the same one that spoils HPLC peaks

Don’t dissolve the sample in a strong solvent and put it onto a weak mobile phase.

Say you’re running hexane and ethyl acetate, and you dissolve the sample in dichloromethane or methanol because that’s what dissolves it. The sample moves down the column in its own strong solvent before the gradient gets to it. The band spreads before the separation starts. You see a fronting or split peak, and the cartridge gets the blame.

Biotage says it plainly: if the crude mixture is only soluble in a strong solvent, dry loading is required, or the separation and purity suffer.

It’s the same error as injecting an HPLC sample in a diluent stronger than the mobile phase. I lost 2 days to that one at the bench years ago, chasing a fronting peak, swapping the column, preparing every sample again. The column was fine. My injection solvent was stronger than my mobile phase.

flowchart TD
    A["Sample ready to load"] --> B{"Does it dissolve in a small<br/>volume of a weak solvent?"}
    B -->|"Yes"| C["Wet load in the<br/>weakest solvent possible"]
    B -->|"No, it needs a<br/>strong solvent"| D["Dry load"]
    D --> E["Mix with sorbent, 1:1 to 1:3<br/>sample to sorbent by weight,<br/>evaporate, load on top"]

SiliCycle’s loading guide says the same: wet load in the minimum volume of the weakest solvent you can, and dry load when the sample isn’t soluble enough in a weak solvent. For dry loading, they suggest 1:1 to 1:3 sample to sorbent by weight, with the solvent removed on a rotary evaporator. Biotage lists silica, C18 and Celite as dry loading media, with Celite releasing everything loaded on it.

If the compound is too polar to hold on silica at all, that’s a different problem. Choosing a column for a polar analyte covers it.

What I check before blaming the cartridge

  1. What was the Rf, and in which solvent system?
  2. What was the ΔCV to the nearest impurity?
  3. How much was loaded, as a percentage of the silica?
  4. What solvent was the sample loaded in?
  5. Wet load or dry load?

Usually the answer is in that list, and a new cartridge wouldn’t have helped.

If you want a second opinion

Send me the TLC, the solvent system, the compound class and roughly how much you need to purify. I’ll tell you the phase, the size and the load I’d use, and whether the cartridge you already have is the right one.

If it is, I’ll tell you that. I’d rather you keep the cartridge and come back next time.

The cartridges I supply are on the flash and preparative columns page.

Sources

  • Biotage, Successful Flash Chromatography, white paper PPS490 v3 (2023): CV = 1/Rf, ΔCV, target Rf 0.15 to 0.4, selectivity before strength, choosing normal or reversed phase, when dry loading is required, dry loading media.
  • SiliCycle, SiliaSep cartridges characteristics: the two silica grades, the loading chart by ΔCV, the Rf to CV table. Read 4 October 2026.
  • SiliCycle, How should I load my compound in my SiliaSep cartridge?: wet and dry loading, the 1:1 to 1:3 ratio.

Both are cartridge makers, and SiliCycle supplies the cartridges I sell. The loading numbers are their guidance, not independent tests.

Checked 4 October 2026.

Common questions

What Rf should I aim for in flash chromatography?
Biotage suggests setting the TLC so the target compound has an Rf between 0.15 and 0.4, in the solvent system you plan to run on the column.
How much sample can I load on a flash cartridge?
It depends on how well separated the compounds are. SiliCycle's loading chart goes from 1% of the silica mass when the difference in column volumes (ΔCV) is 0.1 to 0.6, up to 10% when ΔCV is over 6. On a 12 g cartridge that's 120 mg to 1.2 g.
What is ΔCV?
The number of column volumes between two compounds. Column volumes to elute are about 1/Rf, so a compound at Rf 0.25 elutes at about 4 CV and one at Rf 0.4 at about 2.5 CV. ΔCV is the difference, here 1.5.
When should I dry load a flash column?
When the sample doesn't dissolve well in a weak solvent. Biotage says that if it only dissolves in a strong solvent, dry loading is required, or the separation and purity suffer. SiliCycle suggests 1:1 to 1:3 sample to sorbent by weight.

More guides