Plunger rods, piston seals, check valves, rotor seals, needle seats and inlet filters, for Agilent, Shimadzu, Thermo and Waters pumps and autosamplers. The parts behind a drifting baseline, carryover, and a pressure trace that will not settle.
Cross-referenced
To the OEM part number
Origin and warranty stated
In the quote, before you decide
One Canadian invoice
In CAD, the customs entry is ours
Symptom first
We tell you which part to change
Categories
97 listings
Nothing listed matches that.
The catalogue does not cover everything we can source. Send the part number or the instrument and we will find it and quote it.
Ask for a quote
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat. An acetonitrile-resistant version is available, ask for it if your method runs high acetonitrile.
Check valve cartridge for the Waters 2695 Alliance, cross-referenced to 700000254. A check valve that is not seating gives you a pressure trace that swings with the pump stroke, and the free test is worth running first: sonicate the cartridge in isopropanol…
Plunger assembly for the Waters 2695 Alliance, cross-referenced to WAT270959. The assembled part rather than a bare plunger, which is the safer choice on this pump if you are not confident seating a plunger squarely, because a plunger running even slightly…
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.
Part of the seal-wash circuit behind the high-pressure seal. When it wears, wash solvent leaks or buffer crystals build up behind the head. It is usually changed at the same service as the piston seals.
Part of the seal-wash circuit behind the high-pressure seal. When it wears, wash solvent leaks or buffer crystals build up behind the head. It is usually changed at the same service as the piston seals.
Part of the seal-wash circuit behind the high-pressure seal. When it wears, wash solvent leaks or buffer crystals build up behind the head. It is usually changed at the same service as the piston seals.
Part of the seal-wash circuit behind the high-pressure seal. When it wears, wash solvent leaks or buffer crystals build up behind the head. It is usually changed at the same service as the piston seals.
The wear parts for a routine service of the pump, in one order, so the service is not held up by one missing seal. We send the full contents list with the quote, so you can check it against what your pump needs.
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it. An acetonitrile-resistant version…
Type
Inlet check valve cartridge
Fits
Waters Alliance 2695, 2795, 2796 and the 616 pump (stainless steel heads)
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it. An acetonitrile-resistant version…
Type
Outlet check valve cartridge
Fits
Waters Alliance 2695, 2795, 2796 and the 616 pump (stainless steel heads)
The housing holds the check valve cartridge in the pump head. Order one when the threads or the sealing face are damaged, or when you change to a cartridge that needs its matching housing. Ask us which cartridge it pairs with before you order.
Type
Inlet check valve housing
Fits
Waters Alliance 2695, 2795, 2796 and the 616 pump (stainless steel heads)
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it.
A sapphire piston wears where the seal rides, not evenly along its length. Take it out when you change a seal and look along it under magnification with a light behind it. A ring you can catch a fingernail on means both parts go.
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it. An acetonitrile-resistant version…
The housing holds the check valve cartridge in the pump head. Order one when the threads or the sealing face are damaged, or when you change to a cartridge that needs its matching housing. Ask us which cartridge it pairs with before you order.
A small seal at the outlet check valve. It is cheap, and it is the first thing to check when the outlet valve weeps at the fitting. Sold in packs of 5.
A small seal at the outlet check valve. It is cheap, and it is the first thing to check when the outlet valve weeps at the fitting. Sold in packs of 5.
Type
Outlet check valve adapter seal
Fits
Waters M501, 600, 600E, 610, and extended-flow heads M600 EF, 600E EF, 610 EF, 650 EF
A sapphire piston wears where the seal rides, not evenly along its length. Take it out when you change a seal and look along it under magnification with a light behind it. A ring you can catch a fingernail on means both parts go.
A sapphire piston wears where the seal rides, not evenly along its length. Take it out when you change a seal and look along it under magnification with a light behind it. A ring you can catch a fingernail on means both parts go.
The wear parts for a routine service of the pump, in one order, so the service is not held up by one missing seal. We send the full contents list with the quote, so you can check it against what your pump needs.
The wear parts for a routine service of the pump, in one order, so the service is not held up by one missing seal. We send the full contents list with the quote, so you can check it against what your pump needs.
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.
The housing holds the check valve cartridge in the pump head. Order one when the threads or the sealing face are damaged, or when you change to a cartridge that needs its matching housing. Ask us which cartridge it pairs with before you order.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
A sapphire piston wears where the seal rides, not evenly along its length. Take it out when you change a seal and look along it under magnification with a light behind it. A ring you can catch a fingernail on means both parts go.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
The cartridge is the ball-and-seat insert inside the check valve housing, so it is the part to change when the valve stops sealing and cleaning does not bring it back. Clean the housing and its sealing face before you fit it.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
A sapphire piston wears where the seal rides, not evenly along its length. Take it out when you change a seal and look along it under magnification with a light behind it. A ring you can catch a fingernail on means both parts go.
The housing holds the check valve cartridge in the pump head. Order one when the threads or the sealing face are damaged, or when you change to a cartridge that needs its matching housing. Ask us which cartridge it pairs with before you order.
Sapphire plunger rod for the Agilent 1220, 1260 and 1290, cross-referenced to 5067-4695. These are the later pump heads, and on a 1290 running near its pressure ceiling the plunger and seal both wear faster than the same parts in an 1100 ever did, so the…
Aftermarket active inlet valve cartridge, 400 bar, for the Agilent G1310A, G1311A, G1312A, and the 1260 Infinity capillary and nano pumps, see the full fitment list above, cross-referenced to 5062-8562. A valve that is not seating shows up as pressure that…
Type
Active inlet valve cartridge
Pressure rating
400 bar
Fits
Agilent G1310A, G1311A, G1312A, and the 1260 Infinity capillary and nano pumps. These are found in 1100 and 1200 systems, and in 1260 capillary and nano systems
Does not fit
The B-series pumps (G1310B, G1311B, G1312B) and the G7111 and G7112 pumps are rated to 600 bar and take a different cartridge. Read the module number off the front of the pump, because the system badge does not decide this: a 1200 can carry a 600 bar B-series pump.
Aftermarket active inlet valve cartridge, 600 bar, for the Agilent G1310B, G1311B, G1312B, G7111A, see the full fitment list above, cross-referenced to G1312-60020. A valve that is not seating shows up as pressure that swings with the pump stroke. Clean it…
Type
Active inlet valve cartridge
Pressure rating
600 bar
Fits
Agilent G1310B, G1311B, G1312B, G7111A, G7111B, G7112A, G7112B. These are found in 1200, 1260 Infinity and Infinity II systems
Does not fit
The A-series pumps (G1310A, G1311A, G1312A) and the 1260 capillary and nano pumps are rated to 400 bar and take a different cartridge. Read the module number off the front of the pump, because the system badge does not decide this: a 1260 capillary pump takes the 400 bar cartridge.
PEEK needle seat, 0.17 mm internal diameter, cross-referenced to G1329-87017. The bore is the thing to get right: 0.17 mm is the standard analytical seat, and the narrower 0.12 mm version exists for low dispersion work at small injection volumes. Fitting the…
Needle seat, 0.12 mm internal diameter, cross-referenced to G1329-87012. The narrow bore version, for low dispersion work where the extra volume of a 0.17 mm seat would broaden an early eluting peak. If you run conventional 4.6 mm columns at normal injection…
Needle assembly for the Agilent 1100 and 1200 autosampler, cross-referenced to G1313-87201. The complete assembly rather than the needle alone, which is the easier repair if the needle has been bent and you are not certain the mount survived it. A bent needle…
Injection needle, 900 uL, cross-referenced to G1313-87202. The 900 uL designation refers to the metering device this belongs with, not the injection volume you use day to day, so match it to the injector rather than to your method. The needle tip seals…
Metering piston for the injector, 900 uL, cross-referenced to 5062-8587. A scored piston leaks past its seal, which gives injection volumes that read low and drift through a sequence. Replace the metering seal at the same time. A scored piston is usually what…
Metering piston seal, 900 uL, cross-referenced to 0905-1294. This is the injector seal, not the pump one, and it fails into volume rather than pressure: areas that read low and drift down through a sequence while retention stays exactly where it was.…
Sapphire plunger for the Agilent 1100 and 1200 pump head, cross-referenced to 5063-6586. Sapphire does not wear evenly against a seal. It develops a scored band exactly where the seal rides, and that band is what destroys the next seal you fit. Take the…
Aftermarket gold piston seals for the Agilent 1050, 1100, 1120, 1200, see the full fitment list above, cross-referenced to 0905-1420. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the baseline…
Aftermarket black piston seals for the Agilent 1050, 1100, 1120, 1200, 1220, 1260, cross-referenced to 5063-6589. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the baseline moves with it. If the…
Plunger seal for the metering device, cross-referenced to 5022-2175. This is the injector seal rather than the pump one, so it fails differently: instead of pressure ripple you get injection volumes that read low and areas that drift downward across a…
Aftermarket pump seal for the Agilent 1050, 1100, 1120, 1200, 1220, 1260, G7129A, cross-referenced to 0905-1503. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the baseline moves with it. If the…
Wash seal for the Agilent 1100 and 1200 pump, cross-referenced to 0905-1175. It sits behind the main piston seal and its job is to keep the wash solvent where it belongs, so it fails quietly: no pressure fault, just solvent appearing under the pump head or…
Seal wash kit for the Agilent 1100 and 1200, cross-referenced to G1310-68731. Seal wash is the maintenance that gets switched off and forgotten, and it is the single cheapest thing that extends piston seal life on a pump running buffers. Without it, salt…
Two groove Vespel rotor seal for the Agilent 1100 and 1200 injection valve, cross-referenced to 0100-1853. Count the grooves on the seal you take out before ordering, because the two groove and the standard pattern are not interchangeable and the valve body…
Vespel rotor seal for the Agilent 1100 and 1200 manual injection valve, cross-referenced to 0101-0623 and to the Rheodyne 7125 family. The seal is the wear part in the valve and the stator is not, so unless the stator face is damaged this is the only piece…
Rotor seal for the 600 bar two position six port valve on the Agilent 1200 and 1260, cross-referenced to 0101-1416. The pressure rating is the thing to get right here. A 600 bar valve seal and a 400 bar one look similar and are not the same part, and fitting…
Aftermarket manual injection valve rotor seal for the Agilent 1100, 1200, PerkinElmer 200 Series, Rheodyne 7125, see the full fitment list above, cross-referenced to Agilent 0101-0623. A worn seal leaks between ports in the injection valve, which shows up as…
Aftermarket purge valve with PTFE frit for the Agilent 1100, 1120, 1200, 1260, G1310A/B, G1311A/B/C, G1312A/B/C, G1376A, G2226A, cross-referenced to G1312-60071. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws air on…
Aftermarket glass solvent filter for the Agilent 1050, 1090, 1100, 1200, 1220, 1260, 1290, cross-referenced to 5041-2168. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws air on one channel. It is cheap, and it is…
Aftermarket stainless solvent inlet filter for the Agilent 1050, 1090, 1100, 1200, cross-referenced to 01018-60025. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws air on one channel. It is cheap, and it is blamed on…
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat. An acetonitrile-resistant version is…
Type
Passive inlet check valve assembly
Fits
Agilent 1050, 1100, 1200, 1220 and 1260 Infinity pumps
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat. An acetonitrile-resistant version is…
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
Type
Outlet check valve assembly, 2-piece housing
Fits
Agilent 1050, 1100, 1200, 1220 and 1260 Infinity pumps
The housing holds the check valve cartridge in the pump head. Order one when the threads or the sealing face are damaged, or when you change to a cartridge that needs its matching housing. Ask us which cartridge it pairs with before you order.
Replaces 228-35601-93, old 228-34498-91, 228-35601-92
Sapphire plunger for the LC-20AD and LC-2010 pump, cross-referenced to 228-34498-91. The bare plunger, for when the holder is sound and only the plunger is scored. This is the plunger only, supplied without the holder. Check the holder before ordering: if it…
Sapphire plunger for Shimadzu SIL autosamplers and the LC-2010, cross-referenced to 228-35010-91. This is the autosampler plunger rather than the pump one, and the two are different parts despite looking alike, so check where the fault actually is before…
Plunger holder for the LC-20AD and LC-2010 pump, cross-referenced to 228-35602-91. The holder is what keeps the plunger square to the seal, and it is the part nobody suspects. A worn or distorted holder lets the plunger run slightly off axis, which wears the…
Replaces Shimadzu 228-35146-00, old 228-18745-00, Sciex 445353
Aftermarket plunger seal for the Shimadzu LC-9A, LC-10AD, LC-10ADvp, LC-20AD/AB, see the full fitment list above, cross-referenced to Shimadzu 228-35146-00. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple…
Aftermarket plunger seal for the Shimadzu LC-20AT and LC-10AT pumps, see the full fitment list above, cross-referenced to Shimadzu 228-35145-00. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the…
Type
Plunger seal
Fits
Shimadzu LC-10AS, LC-10AT, LC-10ATvp, LC-20AT, LC-7A, LC-10Ai, LC-6AD, the LC-2010 metering pump, Prominence-i (LC-2030) and Nexera-i (LC-2040)
Replaces Shimadzu 228-32784-91, old 228-24311-01 and 228-31828-00, Sciex 4425151
PTFE diaphragms for the LC-20AD and LC-2010, two per pack, cross-referenced to 228-24311-01. These sit in the degasser rather than the pump head. A perished diaphragm stops the degasser holding vacuum, and the symptom is dissolved gas reaching the pump:…
Vespel needle seal for the LC-2010 and SIL-10ADvp, cross-referenced to 228-33355-04. Vespel is the harder of the two materials Shimadzu uses here and it lasts longer, but it is less forgiving of a needle that is not perfectly aligned, and it dislikes high pH…
PEEK needle seal for the LC-2010 AHT and ACHT, cross-referenced to 228-41540-01. PEEK is the softer of the two options, which makes it more tolerant of slight needle misalignment and better suited to high pH methods where Vespel degrades. The trade is a…
Injection needle for the LC-2010, uncoated, 10 Series, cross-referenced to 228-41024-91. Uncoated is worth noting, because a coated needle is the usual answer to sample carryover with sticky or basic compounds, and this is not that part. A blunted or bent tip…
Aftermarket low pressure valve rotor, PEEK, for the Shimadzu LC-2010, LC-2010 HT, SIL-20 A/C, SIL-20ACHT, SIL-20ACXR, SIL-HT, cross-referenced to Shimadzu 228-36923-00. A worn seal leaks between ports in the injection valve, which shows up as carryover, as…
Aftermarket line filter assembly for the Shimadzu i-Series, LC-10ADvp, LC-10AT, LC-10ATvp, see the full fitment list above, cross-referenced to Shimadzu 228-35871-96. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws…
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
A check valve that is not seating shows up as pressure that swings with the pump stroke, or as retention times that drift. Clean it before replacing it, because a trapped particle gives the same symptom as a worn seat.
Aftermarket check valve cartridge for the Thermo UltiMate 3000 series, DGP-3600BM/RS/SD, HPG-3200RS/SD, HPG-3400RS/SD, see the full fitment list above, cross-referenced to 6041.2301. A valve that is not seating shows up as pressure that swings with the pump…
Aftermarket main piston seals for the Thermo HPG-3200RS, HPG-3400RS, LPG-3400RS, DGP-3600RS, cross-referenced to 6266.0305. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the baseline moves with…
Aftermarket piston seals for the Thermo UltiMate 3000 RS and SD pumps, ISO-3100SD, HPG-3200RS/SD, HPG-3400RS/SD, see the full fitment list above, cross-referenced to 6040.0304. As the seal wears, mobile phase leaks past the piston on the compression stroke,…
Seal wash piston seals for the UltiMate 3000 pump, two per pack, cross-referenced to 6040.0033. These are the rear seals in the wash system rather than the main high pressure seals, and they are the ones that get overlooked. When they go you see wash solvent…
Aftermarket pump head seals for the Thermo UltiMate 3000 series, cross-referenced to 6040.2208. As the seal wears, mobile phase leaks past the piston on the compression stroke, so pressure ripple rises and the baseline moves with it. If the plunger behind it…
Rotor seal kit for the UltiMate 3000 and WPS-3000, cross-referenced to 6840.0012. Carryover on this autosampler is usually blamed on the needle and the wash, and the rotor seal is the quieter cause: worn grooves let a little of the previous sample cross…
Needle seat service kit for the WPS-3000, cross-referenced to 6820.0038A. The needle seat takes the needle on every injection and the seat face is what seals against it, so a scored or coned seat leaks at the moment of injection. The volume you lose is small…
Viper sample needle for the WPS-3000, cross-referenced to 6820.2432. Viper fittings are finger tight by design and that is the part people get wrong: over tightening with a tool deforms the cone and produces exactly the dead volume the fitting exists to…
Metering syringe, 100 uL, for the WPS-3000, cross-referenced to 6822.0002. The syringe meters the injection, so its wear shows up as volume rather than pressure: areas that read low, or precision that falls off at small injection volumes while large ones…
Aftermarket titanium filter frits for the Thermo UltiMate 3000 RS and SD pumps, cross-referenced to 6268.0111. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws air on one channel. It is cheap, and it is blamed on the…
Aftermarket stainless filter frits for the Thermo UltiMate 3000 series, cross-referenced to 6268.011. A blocked filter starves the pump and shows up as pressure ripple, or as a pump that draws air on one channel. It is cheap, and it is blamed on the column…
Piston seals wear with every stroke, and a worn seal shows up as a leak behind the pump head or as pressure that will not hold. Change the seals as a set, and look at the piston under magnification when you do. A scored piston destroys a new seal quickly.