Skip to main content

Independent. Diagnosis first, parts second.

+1 437-505-5995 nazmul@chromkit.ca

Chromkit
Get a quote
Illustration of autosampler vials. Generic drawing, not a photograph of this part. Illustration

This is an aftermarket part, not an original manufacturer part. Instrument names and part numbers identify what it fits and replaces; they remain the trademarks of their owners. The manufacturer, country of origin, rated life and warranty are stated in the quote, before you decide anything. If you need the OEM part, say so and I will quote that instead.

Capillary electrophoresis consumables

Common moulded vials

Replaces OEM part number: A62251

Moulded vials for Beckman CE, cross-referenced to A62251. Moulded glass costs less than drawn tubing and the trade is dimensional consistency, which matters more in CE than in HPLC because the injection depends on the vial sitting at a repeatable height under the capillary. If injection precision has fallen off and the method has not changed, rule the vial batch out before you strip the instrument. Confirm the closure format and the tray fit against your instrument before ordering.

Type
Vials
Fits
Beckman CE
Replaces
A62251
Condition
new
Price
CA$107 before tax and shipping. Confirmed on the quote.

Before you order: if the instrument is misbehaving, tell me the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, I will say so.

How buying works, written for procurement →